6
When the display window shows INSERT TEST DEVICE OR DOUBLE-CLICK BUTTON FOR WALK AWAY MODE, double-click the power
button.
7
If the BD Veritor Plus Analyzer has been configured with barcode ID – Enable Operator ID, it will display SCAN OPERATOR ID. When this
message is displayed, after verifying that the Operator ID to be scanned is correct, the operator should scan his/her identification barcode.
This message is displayed for up to 30 seconds, after which the process will restart with step 6. After a specific Operator Identification has
been recorded, that Operator Identification will be utilized for all subsequent tests until the BD Veritor Plus Analyzer is powered off. On the
next power cycle, the SCAN OPERATOR ID prompt will be repeated during the first test. A test result cannot be generated if the Operator ID
is not input when the Operator ID workflow option is enabled.
BD recommends reading Operator Identification barcodes with the BD Veritor Plus Analyzer positioned at the edge of a flat surface. The
barcode should then be moved toward the barcode window to be scanned. The scanned barcode value will be shown in the next display
window.
8
In the case of Quality Control (QC) Tests, the barcode on the printed label of the swab control supplied with the kit is used as the
Specimen ID.
The display window will now display SCAN SPECIMEN ID. When this message is displayed, the operator should scan the barcode on the
printed label on the swab control supplied with the kit. This message is displayed for up to 30 seconds, after which the process restarts with
step 6. A test result cannot be generated if the swab control barcode is not input.
BD recommends reading Quality Control (QC) barcodes with the BD Veritor Plus Analyzer positioned at the edge of a flat surface. The
barcode should then be moved toward the barcode window to be scanned. The scanned barcode value will be shown in the next display
window.
9
In the case of Quality Control (QC) Tests, the barcode on the printed label of the test device kit box must be scanned.
The display window will show SCAN KIT LOT NUMBER whether the kit information workflow configuration setting is enabled or disabled for
normal operation. When this message is displayed, the operator should scan the barcode on the exterior of the BD Veritor Test Kit box. This
message is displayed for up to 30 seconds, after which the process restarts with step 6. A test result cannot be generated if the Kit Lot
Number is not input.
If the BD Veritor Test Kit box label has two barcodes, scan only the upper barcode which starts with (17).
BD recommends reading Kit Lot Information barcodes with the BD Veritor Plus Analyzer positioned at the edge of a flat surface. The barcode
should then be moved toward the barcode window to be scanned. The scanned barcode value will be shown in the next display window.
10
The display window will now show ADD CONTROL TO TEST DEVICE AND INSERT IMMEDIATELY. This message is displayed for up to
3 minutes, after which the process restarts with step 6. Apply the prepared control to the test device sample well as instructed in the test
device kit package insert and immediately insert the test device into the slot on the right side of the Analyzer, aligning the insertion arrow on
the test device with the arrow above the slot. Insert the device fully until it stops. A distinctive "click" will be noted when the device is properly
aligned in the Analyzer. During this process the test device must remain horizontal to prevent spilling the specimen out of the sample well.
11
The display window will now show DO NOT DISTURB TEST IN PROGRESS. The incubation time is determined based on the test device
barcode. The incubation time remaining is shown on the display. Do not disturb the test device or Analyzer during the incubation period. Do
not remove the test device; doing so will cause the test to abort.
12
After the incubation is complete, the BD Veritor Plus Analyzer will progress through two processing steps: a reading step, followed by an
analyzing step. The display will show the remaining time for each step as they are performed. Do not touch the Analyzer or remove the test
device during this time.
13
When the analysis is complete, the test result will be displayed with the control swab identification, the name of the test, a result, and an
overall "QC PASS" or "QC FAIL" indication. Refer to the package insert for the BD Veritor Test Device for specific result interpretations. If a
printer is connected to the unit and powered on, the test result will automatically be sent to the printer.
14
Verify that the test type and QC swab control are correct.
15
Once the test type and swab control are verified and the result noted, remove the test device by pulling it out. The display will show INSERT
TEST DEVICE OR DOUBLE-CLICK BUTTON FOR WALK AWAY MODE to indicate the Analyzer is ready to perform another test.
16
If an InfoSync Module is installed, the ENVELOPE symbol will appear to indicate that results are being transmitted over the secure cellular
connection. The symbol will disappear after the results have been transmitted. In the event that the BD Veritor Plus Analyzer does not detect
adequate cellular network strength, it will queue all results to be transmitted and continuously attempt to transmit the results while it is
powered on. If the Analyzer is powered off while the ENVELOPE symbol is still present, the Analyzer will queue the result and transmit it the
next time that it is powered up.
17
To initiate a new test, repeat steps 6–16 above, or turn off the power by depressing the front panel power button for at least half a second
and releasing it. If the Analyzer is left unattended for 60 minutes, the BD Veritor Plus Analyzer will automatically shut off and the test result
will not be retained on the screen.
17