Seca 780 Manual Del Usuario página 9

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 15
What do I do if ...
... the scales fl ash because they have been
loaded for over 1.5 min?
Briefl y take the load off the scales until
they display
.
... the display fl ashes because the scales
have been exposed to severe changes in
temperature?
Put the scales somewhere else or allow
them to adjust to the new
temperature.
...
appears in the display?
Batteries will soon be running out,
change batteries.
...
appears in the display?
Change batteries
... a number, e.g. 12, appears in the right-
hand side of the display?
You have inadvertently pressed seve-
ral keys at once and put the scale in
"service mode". Do not press any more
keys and turn off the scales using the
green Start key. You can then start the
scales quite normally again.
... the display stays dark after being
switched on?
You've almost certainly forgotten to put
batteries in.
... no weight display appears?
Check whether the load lever is
correctly suspended from the
connecting rod.
... with the scales unladen,
appears in the display instead of exactly
?
Re-set scales to zero: start again using
the green Start key – the display will fi rst
show
automatically jump to zero.
...
appears in the display after the
scales are started?
Too great a load was placed on the
scales when starting them. Take load
off scales and re-start using the green
Start key.
...
appears in the display?
The scales are being operated in the
incorrect temperature range. Ensure
that the scales are operated between 0
°C and 40 °C.
...
appears in the display?
Maximum load has been exceeded.
Servicing
Your seca scale leaves the factory with an
accuracy better than ± 0,4 / 0,2 %. To
ensure that this degree of accuracy is
maintained, the product must be carefully
set up and regulary serviced. Depending on
frequency of use, we recommend servicing
at intervals of 3 to 5 years. Please note that
a qualifi ed servicing agent should be used
for this. If you are in doubt, seca Customer
Service will give you more
information.
or
and then
Changing the batteries
When
or
appear in the display,
you should change the batteries. You need
four type AA 1.5 V batteries.
• Carefully turn over the scale so that the
base of the scale is accessible.
• Push the cover of the battery compart-
ment towards "Batterie" and open the cover.
• Take out the holder and batteries.
• Remove the used batteries and insert
new batteries in the holder.
Check that the polarity is correct.
• Place the holder and batteries in the
battery compartment.
• Close the cover.
• Then turn the scale back up the right way.