Installing the push-in
5.3 Set up device
82 •
ATTENTION!
Incorrect measurements due to installation
scale
errors
If the push-in scale is installed incorrectly the
head slide cannot determine any usable
readings.
− Align the push-in scale so that the printing
− Align the push-in scale so that the area which
− Ensure that the push-in scale latches into
Push the push-on scale into the groove in the side
of the column elements until the scale latches into
position under the end cap.
ATTENTION!
Incorrect measurement with force shunt
Correct measurements cannot be taken if the
patient's weight is not fully positioned on the
weighing area.
− Set up the device so that it only makes
− Set up the device so that the weighing area
− While you are weighing, ensure that the
1. Place the device on a flat, stable surface.
remains visible when the scale is pushed into
the groove.
shows the product number forms the bottom
end.
position under the end cap.
contact with the ground via the foot screws.
does not touch any other objects.
patient does not touch the upright of the
device.