Seca 285 Manual De Instrucciones Y Garantía página 80

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 253
Install the
multifunctional display
80 •
The multifunctional display is hooked into the groove of
the column element and fixed with a clamping screw.
The installation height is freely selectable.
1. Insert the T-slot nut of the multifunctional display
into the groove of the second column element.
2. Tighten the two hexagon socket head screws to fix
the multifunctional display in the required position.
3. Guide the connection cable of the multifunctional
display into the groove of the column element up to
the base of the device.
4. Pull the connection cable through the hole in the
base of the device.
ATTENTION!
Malfunction due to installation errors.
If the cables are installed so that they are
subjected to large mechanical stress, faulty
displays and failure of the display can result.
− Lay all cables so that they are not bent too
tight and the plugs are not kinked.
5. Wind the free end of the connection cable onto the
cable winding in the base.
6. Plug the connection cable plug into the matching
socket of the weighing platform.
loading

Este manual también es adecuado para:

285dp