Maintenance
Maintenance Inspection and Safety Management
The HBP-1120 must be maintained to ensure functionality and to secure the
safety of patients and operators.
Daily checks and maintenance should be performed by the operator.In
addition, qualified personnel are necessary to maintain the performance
and the safety, and to conduct periodic inspections. We recommend that the
verification test be performed at least once a year.
Cleaning of the Device
Cleaning and disinfecting should be performed in accordance with your
facility's infection control practice.
Wipe with a cloth that has been moistened with isopropyl alcohol diluted to
50 v/v%, or ethyl alcohol (disinfection alcohol) diluted to 80 v/v% or less and
wrung out.
Do not wipe the Power connector or allow it to become wet.
Use a moistened cotton bud to remove dust that has accumulated on the vent
ports.
The device requires no routine service other than cleaning, and visually
checking the cuffs, tubing, etc.
Caution
• Do not sterilize by autoclave or gas sterilization (EOG, formaldehyde gas,
high-concentration ozone, etc.).
• If using an antiseptic solution for cleaning, follow the instructions of the
manufacturer.
Cuff Care
Wipe clean on the surface of the cuff with a cloth moistened with a 70 v/v%
dilution of isopropyl alcohol, or a 80 v/v% or less dilution of disinfection
ethanol (ethyl alcohol).
Do not allow any liquids inside the cuff. If a liquid gets in the cuff, dry the
inside well.