8. Spare part kit
8.3
Mounting
EN
16
Secure the clamping jaw in the vice as shown, so that
this cannot come loose.
Slide the pressure sensor between the clamping jaw,
using the groove milled into the process connection.
Loosen the process connection using the torque
spanner.
WIKA operating instructions pressure sensor, model HP-2
Dry the baffle plate incl. the
O-ring using compressed air.
For access, there is a bore
on the process connection.
Baffle plate
DIPS
EPC