Mettler Toledo SevenGo Duo pro SG78 Instrucciones De Manejo página 37

Medidor de ph/orp/ ion/conductividad
Ocultar thumbs Ver también para SevenGo Duo pro SG78:
Idiomas disponibles

Idiomas disponibles

Limits of error ion
Limits of error conductivity
pH calibration
Isopotential point
pH Calibration buffer
Conductivity calibration
standard
Power requirements
Size/weight
Display
pH input
Conductivity input
pH T input
IP rating
Ambient conditions
Materials
SevenGo Duo pro™
± 0.5% (this limit only applies for
meter)
Conductivity
TDS
Salinity
Resistivity
Temperature
Up to 5 points
pH 7.00
7 predefined groups
5 predefined standards
Ratings
Batteries
220 x 90x 45 mm
342 g
Liquid crystal
BNC,
impedance >
3 * 10e+12 Ω
7-Pin LTW plug
RCA (Cinch), NTC 30kΩ
IP67 with and without electrode
Temperature
Relative humidity
Installation category
Pollution degree
Altitude
Housing
Window
Keypad
±0.5 % of measured value
±0.5 % of measured value
±0.5 % of measured value
±0.5 % of measured value
±0.1 °C
1 user-defined group of 5 buffers
1 user-defined standard
6 V DC, 70 mA
4 x AA/LR6 1.5 V or NiMH 1.2 V
rechargeable
5...40°C
5%...80% (non-condensing)
II
2
Up to 2000 m above sea level
ABS/PC reinforced
Polymethyl methacrylate (PMMA)
Silicone rubber
Specifications​ ​ 33
loading