Function; Limitations Of Use; Operating Exclusions - BRAND accu-jet S Instrucciones De Manejo

Ocultar thumbs Ver también para accu-jet S:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 107
12. Protect the power adapter from moisture and use it only with this in-
strument.
13. Stop using the power adapter if the connection cable is damaged.
14. Only authorized service technicians are permitted to carry out re-
pairs and internal maintenance on the instrument.

2.1 Function

The device serves as an aid for pipetting liquids with graduated and volu-
metric pipettes made of glass or plastic in the volume range from 0.1 ml
to 200 ml and with a suction tube outer diameter < 9.2 mm. When the de-
vice is handled correctly, the pipetted liquid comes into contact with the
pipette only.

2.2 Limitations of use

This device is intended for pipetting liquids within the following physical
limitations:
• from 10 to 40°C (from 50 to 104°F) (device and reagent)
• vapor pressure up to 500 mbar. Absorb slowly above 300 mbar in or-
der to prevent the liquid from boiling.
• Density up to 9 g/cm

2.3 Operating exclusions

The instrument must not be used for liquids with vapors that can cause
corrosion or attack the materials silicone or EPDM.
The instrument is not suitable for use with Pasteur pipettes.
997418
3
Operating manual
45
Tabla de contenido
loading

Tabla de contenido