Product description
Vacuum system SC820G / SC840G
Design of the vacuum system
The collection flask (5) catches particles and droplets at the
inlet of the pump which have been suctioned out of the recipi-
ent contrary to the requirements of the pump. The collection
flask is coated (implosion protection) and mounted on the sep-
arator 9 with a flask clamp (3).
The high-performance condenser (2) at the pump outlet recov-
ers solvents back from the transferred gas instead of letting
them escape into the environment or into the fume cupboard.
The condenser is lined for thermal insulation and as protection
against bursting.
The solvents that are precipitated in the condenser are col-
lected in the collection flask (4), which is coated (bursting pro-
tection). A flask clamp (3) fixes the collection flask to the con-
denser flange. A circulating cooler or running cold water (or
other cooling medium) cools the high-performance condenser
to condensation temperature.
Translation of Originial Operating Instruction, KNF 339416-339418 02/23