Measuring Temperature
This measurement function is applicable for the U1194A model only.
Set up your clamp meter to measure temperature as shown in
the test points, and read the display.
Do not connect the thermocouple to electrically live circuits. Doing so will
WARNING
potentially cause fire or electric shock.
Do not bend the thermocouple leads at sharp angles. Repeated bending
CAUTION
over a period of time can break the leads.
– The clamp meter uses the type-K thermocouple probe (included in the
NOTE
– Shorting the
Figure 2-19
Keysight U1190A Series User's Guide
standard shipped items of a U1194A model) for measuring temperature.
terminal to the
the clamp meter's terminals.
Temperature display
Making Measurements
terminal will display the temperature at
2
Figure
2-20. Probe
59