Follow the steps below to change the timer period or completely disable the
flashlight timer.
1 Press and hold
2 Press
3 Press
4 Press
5 Cycle the clamp meter's power to exit the Setup menu.
Alerts and warnings during measurement
Voltage alert
For your own safety, please do not ignore the voltage alert. When the clamp
WARNING
meter cautions you with a voltage alert, you are advised to take note of the
existence of high voltage and pay closer attention when performing
measurements.
Your clamp meter provides a voltage alert for voltage measurements. The clamp
meter starts beeping periodically once the measured voltage exceeds the alert
value (regardless of polarity) set.
Hazardous voltage indication
The clamp meter will also display the hazardous voltage ( ) symbol as an early
precaution when the measured voltage is equal to or greater than 30 V in all
voltage measurement modes.
Keysight U1190A Series User's Guide
while powering on the clamp meter to enter the Setup
menu.
twice. The flashlight (t #) timer period is the third Setup item shown.
to make changes to the t # value.
Press
or
to change the timer period (from t 01 to t 99 seconds) or to
completely disable the flashlight timer function (toFF).
to save the changes, or press
proceed to the next Setup item.