3.3.4. Calibration
Turn on the scale, press the [#] key during initialization and the scale will enter the
calibration mode. Calibration steps below:
STEP
Press [TARE] to select the scale
1
Press [TARE] to select the
2
DECIMAL POINT
Set the full range of the scale
3
Zero point calibration:
4
Press [ #] when the scale is
stabilized
Full scale calibration: press [ #]
when the input value is the
5
same as the loaded weight and
the weight is stabilized
6
3.4.
Cleaning and maintenance
a)
Pull out the mains plug before each cleaning and when the unit is not in use.
b)
Use only non-corrosive cleaning agents for cleaning the surfaces.
c)
Store the unit in a dry and cool place protected from moisture and direct
sunlight.
d)
Do not spray the unit with a stream of water or immerse it in water.
e)
Make sure that no water enters through the ventilation openings in the casing.
f)
Clean the ventilation openings with a brush and compressed air.
g)
Perform regular inspections of the unit checking technical fitness and any
damages.
h)
Do not leave the batteries in the unit when it will not be used for a long time.
i)
Use a soft, damp cloth for cleaning.
j)
Do not use sharp and/or metal objects (e.g. wire brush or metal spatula) as
they may damage the surface of the material from which the device is made.
k)
Do not clean the product with acidic substances, medical agents, thinners, fuel,
oils or other chemicals. It may cause damage to the device.
INSTRUCTIONS FOR THE SAFE DISPOSAL OF BATTERIES.
The devices are equipped with 6 V, 1300 mAh batteries.
Remove used batteries from the unit in the same way you put them in.
Return the batteries to a unit responsible for their disposal.
ACTION
DISPLAY
[d X]
Select the scale (1/2/5/10/20/50), press
[ #] to confirm
Example: 20
[PX]
Select the decimal place: 0 ~ 3, press [
#] to confirm
Example: 3
[FULL]
Press [TARE] to change the digit value;
press [ZERO] to select a digit;
Press [ #] for to confirm the set range
[nOLOAD]
Make sure there is no load on the scale
[AdLOAD]
When entering the loaded weight, press
[TARE] to select the digit value. Press
[ZERO] to select a digit. When the input
value is the same as the load weight,
press [#] when the weight is stabilized
[End]
End of calibration