• If the version in the upper right corner has not changed, the SW update failed. Repeat
the ADCC SW update procedure. If this update fails, please contact your local Alere
Afinion supplier.
• Power must not be disconnected during SW update. Loss of power during SW
update might lead to ADCC failure.
• During SW update of ADCC, the unit should be disconnected from the Alere Afinion
AS100 Analyzer. Tests performed on an Alere Afinion AS100 Analyzer connected to an
ADCC during SW update might lead to loss of result transfer to HIS/LIS system.
8 ADCC TROUBLESHOOTING
8.1 LED status indicator signals
Symbols
Situations
Power off
Power on
Power on, no HIS/LIS communication.
ASTM/HL7: Wait 1 min. Check ADCC
configuration and Ethernet connection.
POCT1-A: See section 11.
Power on, no Alere Afinion AS100 Analyzer
communication Analyzer in Automatic start
up procedure. Wait 1 min after Analyzer has
performed Automatic start-up procedure
Alere Afinion AS100 Analyzer connection
established
ASTM/HL7: HIS/LIS connection established
POCT1-A: Data transfer in progress
Connection succeeded, HIS/LIS and
Alere Afinion AS100 Analyzer connection
established
Alere Afinion AS100 Analyzer
communication failed. Wait 1 min. Analyzer
in self-test. Wait 1 min. after Automatic start-
up procedure is performed
HIS/LIS communication fatal error
Unrecoverable error not related to port
communication
SW upgrade is finalizing
Factory reset button activated
If the results are not transferred but the LED lights are not displaying any of the errors
above, please reboot ADCC. If the error persists, please contact your local Alere Afinion
supplier.
For US customers a toll free line for technical support is available
EN
US
8
Alere Afinion
Power LED
(Green)
OFF
Steady
Steady
Steady
Steady
Steady
Steady
Steady
Steady
Steady
Steady
Single flash
Data Connectivity Converter
™
IOIOI
RS232 LED
Ethernet LED
(Yellow)
(Yellow)
OFF
OFF
Any state
Any state
Any state
OFF
OFF
Any state
Steady
Any state
Any state
Steady
Steady
Steady
Flashing
Any State
Any State
Flashing
OFF
OFF
LED is lit in sequence during upgrade. SW
upgrade is successfully performed when LED
flashes three times simultaneously
Single flash
Single flash
AS100
Warning LED
(Red)
OFF
Any state
Any state
Any state
Any state
Any state
OFF
Flashing
Flashing
Flashing
Single flash