Greenlee 93-97/00220 Manual De Instrucciones página 18

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

MAINTENANCE
Do not attempt to repair this test instrument. It contains no user-
serviceable parts.
Failure to observe this precaution can result in injury and can damage
the instrument.
General
• Keep unit clean and dry.
• Wipe clean with damp cloth. Let
dry completely before using.
Battery Condition
Battery level is low when -
appears on display.
Battery Replacement (Fig. 2)
1. Disconnect the test leads from
item being tested and turn meter
off. Unplug leads from front
terminals. Remove unit from protective holster.
2. Position meter face down on clean, flat surface. Remove the three screws
from case back.
3. Lift the bottom end of the case back until it gently unsnaps from the case
front at the top end near the display.
4. Lift the battery holder out of the meter and carefully remove the batteries.
5. Install two new AAA (1.5V) alkaline batteries. Be sure the leads will not be
pinched when case is reassembled.
6. Replace case back. Make sure all gaskets are properly seated and the
two snaps at the top are engaged. Replace the screws and tighten firmly.
Do not overtighten.
18
All manuals and user guides at all-guides.com
+
Batteries (2)
Battery
Holder
Case Back
Case Front
Fig. 2
Tabla de contenido
loading

Tabla de contenido