Electromagnetic compatibility
Table 1:
Guidance and manufacturer's declaration - electromagnetic emissions
The Device is intended for use in the electromagnetic environment specified below. The customer or the user of the
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Table 2:
Guidance and manufacturer's declaration – electromagnetic immunity
The Device is intended for use in the electromagnetic environment specified below. The customer or the user of the
Immunity test
Electrostatic di-
±2kV ±4kV ±6kV ±8kV
scharge (ESD)
±15kV
lEC 61000-4-2
Electrical fast
±1kV, ±2kV, 100 kHz
transient/burst
repetition
IEC 61000-4-4
frequency
Surge
±0.5kV, ±1kV (Line to line)
lEC 61000-4-5
±0.5kV, ±1kV, ±2kV
(Line to
Ground)
Voltage dips,
0% U
short interrupti-
for 0.5 cycle
ons and voltage
variations on
0% U
power supply
for 1 cycle
input lines
70% U
Voltage dips,
for 25/30 cycles
short interrupti-
ons and voltage
0% UT
variations on
for 250/300 cycles
power supply
input lines
IEC 61000-4-11
Device should assure that it is used in such an environment
Compliance
Group 1
Class B
Class A
Complies
Device should assure that it is used in such an environment
IEC 60601
test level
±8 kV contact
±2kV ±4kV ±8kV ±15kV
air
Power supply lines ±2 kV
Line to line: ±0.5kV, ±1kV
Line to Ground: ±0.5kV,
±1kV, ±2kV
0% U
T
for 0.5 cycle
0% U
T
for 1 cycle
70% U
T
for 25/30 cycles
0% U
for 250/300 cycles
Electromagnetic environment – guidance
The Device uses RF energy only for its internal func-
tion. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
The Device is suitable for use in all establishments In-
cluding domestic and those directly connected to the
public low-voltage power supply network that supplies
buildings used for domestic purposes.
Compliance
level
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %
Mains power quality should be that of a
typical commercial or hospital environ-
ment.
Mains power quality should be that of a
typical commercial or hospital environ-
ment.
Mains power quality should be that of a
T
typical commercial or hospital environ-
ment.
T
T
Mains power quality should be that of a
typical commercial or hospital environ-
ment. If the user of the model TL55 re-
T
quire continued operation during power
mains interruptions, it is recommended
that the model 168 DAYLUX daylight
therapy lamp be powered from an unin-
terruptible power supply or a battery.
138
Electromagnetic environment –
guidance