EN
User manual - English
Congratulations on acquiring a new Mach LED 3 operating light.
Please read these Instructions for use very carefully.
1. Instructions for safe use
1.1
Intended user
The Mach LED 3 is a Class I medical product and may only be oper-
ated by trained medical staff.
1.2
Information and obligation of the user to check the product
Pay attention to the instructions for use when handling the lamp.
These Instructions for use are part of the product and must therefore
be stored in a place close to the product in order for the safety in-
structions and important information to be consulted at any time.
Make sure that the lamp is in satisfactory working order before every
use. If there is obvious damage, unusual operating conditions, etc.,
the lamp must not be used.
1.3
Availability of the instructions for use
These instructions for use and a detailed handbook with detailed in-
formation on how to deal with a fault, a list of accessoires and further
tips on how to use the light can be found online at the following link:
https://dr-mach.de/login/mach-led-3.html
1.4
Intended use / contra-indications
The Mach LED 3 operating light is designed to illuminate an operating
site in medical facilities (e.g. in a laboratory, in hospitals or doctor's
practice) with focussed, low-glare, shadow-free light. It enables the
user to perform a diagnosis or carry out medical interventions. The
Mach LED 3 light is an operating light that is not fail-safe when used
as a single light. It is not intended for use in areas where explosions
are likely although it is permissible to use it in the vicinity of HF sur-
gical equipment.
Permanent illumination of the open human eye should be avoided
when illuminating the face area.
59181001
1.5
Technical data
Class of protection
I
IP protection class
IP 42
Input voltage
100-240 V AC, 50/60 Hz
(power supply)
Input voltage
24-30 V DC
(light body)
Power consumption
26 W (without camera)
(Mach LED 3SC – 140.000 lux)
36 W (with camera)
Power consumption
30 W (without camera)
(Mach LED 3SC – 160.000 lux)
40 W (with camera)
Power consumption
60 W (without camera)
(Mach LED 3 Smart)
65 W (with camera)
Current
1.1 A max.(without cam-
(Mach LED 3SC – 140.000 lux)
era)
1.5 A max.(with camera)
Current
1.3 A max.(without cam-
(Mach LED 3SC -160.000 lux)
era)
1.7 A max.(with camera)
Current
2.5 A max.(without cam-
(Mach LED 3 Smart)
era)
2.7 A max.(with camera)
Operating time
Continuous operation pos-
sible
(1)
Expected life
10 years
1
At the end of the expected (designed) service life, the lamp must be ser-
viced more frequently for safe operation (see the manual for details).
1.6
Lighting technical data
Mach LED
3SC
Central light intensity
140,000 lux
(distance 1 m)
(optional
160,000 lux)
Light field diameter d10
158 mm
1
165 mm
Light field diameter d50
92 mm
1
96 mm
Residual light intensity
66 %
1
(one shadower)
62 %
Residual light intensity
46 %
1
(two shadowers)
48 %
Residual light intensity
100 %
(normed tube)
5 / 70
Residual light intensity
(normed tube, one shad-
ower)
Residual light intensity
(normed tube, two shad-
owers)
Illumination depth (20%)
Illumination depth (60%)
Radiation intensity in the
field
(distance 1 m)
Max. radiation intensity
in the field
(distance 0.86 m)
Max. radiation intensity
in the field
(distance 0.89 m)
1
Mach LED 3SC with 160,000 lux (optional)
For a complete overview of technical and lighting specifications, refer
to the manual.
1.7
Installation/Maintenance/Repair
The lamp may only be installed, maintained or repaired by the man-
ufacturer or by specially trained staff. Maintenance must be carried
out at least every two years
1.8
Environmental conditions for operation
Ambient temperature:
Mach LED 3
Relative air humidity:
Smart
Air pressure.
140,000 lux
1.9
Reporting obligations
170 mm
Every serious incident which has occurred in connection with the
product must be reported to the manufacturer and the competent au-
92 mm
thority.
61 %
2. Images on the device
49 %
100 %
Mach LED 3
66 %
61 %
1
62 %
46 %
49 %
1
48 %
998 mm
1200 mm
1
1045 mm
476 mm
530 mm
1
520 mm
2
504 W/m
499 W/m2
1
595 W/m2
590 W/m2
-
1
666 W/m2
537 W/m2
+5°C to +40°C
30% to 75% RH
700h Pat o 1060h Pa
This symbol tells the user to observe the
instructions for use.
V2