Inlet Screen Installation - Varian TV 301 Manual De Instrucciones

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 12
N.
Read/
Data
Description
Write
Type
400
R
A
CRC EPROM
(QE)
402
R
A
CRC Param. (PA)
404
R
A
CRC Parameter
structure
500
Reserved to Varian service
503
R/W
N
RS 485 address
504
R/W
L
Serial type select
NOTES
1. Automatic means that when the controller
stops, the vent valve is opened with a delay de-
fined by window n. 126; when the controller
starts, the vent valve is immediately closed.
2. On command means that the vent valve is
opened or closed by means of window n. 122.
3. These commands remove the pump motor
supply for a short time. If they are used con-
tinuously, they can cause the pump stopping.
7
6
5
4
TOO HIGH LOAD
SHORT CIRCUIT
OVERVOLTAGE
AUX FAIL
Window N. 206 Bit Description
All manuals and user guides at all-guides.com
Admitted
Values
QE8XXXX
(where "XXXX" are
variable)
PA8XXXX
(where "XXXX" are
variable)
"XXXX"
0 to 31
(default = 0)
0 = RS 232
1 = RS 485
(default = 0)
3
2
1
0
NO CONNECTION
PUMP OVERTEMP.
CONTROLL. OVERTEMP.
POWER FAIL
36
TECHNICAL INFORMATION

INLET SCREEN INSTALLATION

The inlet screens mod. 969-9302 and 969-9303
prevent the blades of the pump from being dam-
aged by debris greater than 0.7 mm diameter.
The inlet screen, however, does reduce the pump-
ing speed by about 10%.
The inlet screen is fitted in the upper part of the
pump, as shown in the figure.
The screen can be mounted on each pump.
87-900-946-01(B)