Serial Cable Installation; Tv 301 Controller Installation; Bottom Mounting - Varian TV 301 Manual De Instrucciones

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 12
and then screw the gas purge valve (with a torque
of 2.5 Nm) as shown in the following figure.

SERIAL CABLE INSTALLATION

The supplied serial cable must be installed when
the TV 301 Navigator or the optional vent valve
have to be controlled by means of a remote com-
puter.
The cable is installed fixing the 9 pin D-type con-
nector into the P2 serial connector as shown in the
following figure.
All manuals and user guides at all-guides.com
42
TECHNICAL INFORMATION

TV 301 CONTROLLER INSTALLATION

The controller can be mounted in two position:
bottom mounting (as per the complete system)
side mounting.
To install the controller execute the following pro-
cedures.

Bottom mounting

See the following figure.
1. Turn the pump upside-down;
2. Screw the 4 fixing studs (provided with the ac-
cessories bag) in the holes on the bottom of
pump;
3. Place the controller on the studs aligning the 4
rubber holes on the top of the controller case;
4. Press the controller towards the pump;
5. Turn the pump again;
6. Plug the line cord and connect the pump cable
to the pump;
7. Connect the mating connector with the jumper
on the interlock signal to start the pump.
MOUNTING 1
87-900-946-01(B)
Tabla de contenido
loading

Tabla de contenido