If these instructions are not observed, it may result in
malfunction or damage to the equipment.
Tips and advice that make the work easier.
2. Installing the product
2.1 Location
Install the product in a location that meets the following
requirements:
•
Place the product in a flood-safe place.
•
Make sure that the ambient temperature is within the limits.
•
Install the product as close as possible to the connected
pumps, sensors, and accessories.
•
The product must be protected from direct sunlight.
•
The product must be easily accessible.
A
B
Type
Dimensions [mm] (inches)
A
278
Plastic cabinet
(10.9)
500
Metal cabinet
(19.7)
500
Metal cabinet
(19.7)
A
B
6
E
B
254
(10)
455
(17.9)
455
(17.9)
E
•
We recommend that you install the product in a protective shed
or enclosure to avoid direct sunlight and rain.
•
Indoor installation: The product must be installed in a well-
ventilated room to ensure cooling of its components.
2.2 Mechanical installation
2.2.1 Installing the control unit
The product is designed to be mounted on a flat and vertical
surface. The cable glands must face downwards.
To ensure sufficient ventilation, allow at least 150 mm
(5.9") of space on either side of the cabinet.
1.
Drill holes in the surface. Use the measurements below
depending on the type of cabinet. See the product's data
booklet for more information about cabinet types.
2.
Insert wall plugs, if applicable.
3.
Fit the four screws in the mounting holes and cross-tighten the
screws.
C
D
378
354
(14.9)
(13.9)
600
525
(23.6)
(20.7)
700
625
(27.6)
(23.6)
E
180
(7)
250
(9.8)
250
(9.8)