Do NOT turn the power on until instructed in Step 11.
Step 8: Install a Printer (Optional)
Install a USB point-of-service (POS) printer, (such as a Citizen
CD-5500A or Epson TM-722OD).
1. Connect the printer cable's USB host connector into one of
the USB host ports on the Analyzer.
2. Connect the cable's USB device connector to the printer.
3. Configure the printer to the Analyzer later in Step 12.
Step 9: Install the iOS System
(When connecting the Portable or OnLine Analyzer for online
measurements.)
Note:
I
f using a Sievers Autosampler, DO NOT proceed with this sec-
tion. Instead, refer to the Sievers Autosampler Installation Guide.
1. Ensure that the power to the Analyzer
is off and disconnected.
2. Verify that the water source is
disabled.
3. Connect the 1/4" PFA inlet tubing (w/
in-line filter) into the sample inlet (A)
on the iOS System. Secure nut with
9/16" open-end wrench. Do NOT
over-tighten.
4. For a two-stream option (M5310 C
On-Line Only), connect the other inlet
tubing into the second sample inlet.
5. Slide one end of the waste tubing
over the waste outlet fitting (B) and
position the tubing for a gravity drain.
6. Ensure that the stainless steel tubing
is connected to the Sample Inlet
and establish a water flow to the iOS
System.
7. Measure the flow rate and adjust to
between 50-300 mL/min using the
flow rate adjustment screw (C).
Step 10: Install the Waste Outlet
(When configuring the Laboratory for grab or Autosampler use.)
1. Using a 7/16" open-end wrench, disconnect the nut that con-
nects to the Waste port.
2. Attach the 1/8" OD plastic waste tubing to the Waste port,
then use the 7/16" open-end wrench to tighten securely.
3. Position the hose for a gravity drain of the waste water.
Step 11: Prime the DI Water Pump
Note:
The DI water reservoir was filled in Step 5 and the DI pump is now
ready for priming.
1. Ensure that the resin cartridge is in the Prime position (Step 5).
2. Press the power switch to On and let the Analyzer run 180 min.
3. After 180 minutes, on the Home screen, shut down the
system and turn off power to the Analyzer.
Warning:
operating system, always FIRST shut down the system on the
Analyzer's Home screen before pushing the Analyzer's power
switch to the Off position.
1. View the reservoir water level and verify that the float is
positioned JUST BELOW the top of the
reservoir. (The color of the resin in the
cartridge should be black.)
2. If the water level is now lower due to
the circulation process, fill the DI water
reservoir as needed. There is no need to
prime again.
3. Install the DI resin cartridge into the RUN
position. Ensure that the ends are secure.
4. Close and latch the Analyzer's door
(On-Line model) or replace the panel
(Laboratory and Portable models).
C
Step 12: Configure Basic Analyzer Settings
Before placing the Analyzer into service, record the installation
of the reagents on the Consumables tab of the Maintenance
screen. In the appropriate Acid or Oxidizer section, press R
A
Use the Analyzer screen to configure various basic settings,
B
including optional security (Password Protection or DataGuard)
using the touchscreen display. Most of these settings will not
need to change unless moving the Analyzer or reconfiguring
the operational environment.
Step 13: Flush the Analyzer
Use these steps to flush the reagent syringes.
To also rinse the Analyzer's sample fluidics (optional for low
TOC applications), refer to the Analyzer's Operation and
Maintenance Manual for instructions.
To flush the reagent syringes
iOS
1. Connect to either a continuous DI water source or a large
flask (1000-2000 mL) of DI water.
2. Power on the Analyzer.
3. On the Maintenance screen, select the Advanced tab.
4. Press Syringe Flush and follow the on-screen prompts.
5. Shut down the system and turn the power switch to Off.
Step 14: Install a Sievers Autosampler
(Laboratory and Portable TOC Analyzers Only)
If planning to use a Sievers Autosampler with the Analyzer,
install it now. Refer to the Sievers Autosampler Installation
Guide (available for download from our website) for complete
installation instructions.
quick start guide
To avoid the possibility of corrupting the Analyzer's
M-Series Quick Start Guide DQS 77000-02 MUL Rev A
RUN Position
.
enew