Display
Meaning
Battery depleted
BAT
Measuring range
ERR.1
exceeded
Measuring range is
ERR.2
undercut
System error
ERR.7
32 / 36
TD300 series
Possible causes
Battery depleted
Measurement too
high
Measuring probe or
product defect
Measurement too
low
Measuring probe or
product defect
Error in the product Switch the product on/off
Remedy
Replace battery
The measurement is above
the permissible range
Check measuring probe
Send in for repair
The measurement is below
the permissible range
Check measuring probe
Send in for repair
Replace batteries
Send in for repair
B-H86.0.0X.DKZ01-1.0