Cleaning And Disinfecting - FujiFilm SonoSite Edge Suplemento Del Manual De Usuario

Ocultar thumbs Ver también para SonoSite Edge:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 43
2
PG mean = sum(4v
)/N
where:
v =peak velocity at interval n
N = the number of intervals in the Riemann sum
Qp/Qs
Qp/Qs = SV Qp site/SV Qs site = RVOT SV/LVOT SVOh, J.K., J.B. Seward, and A.J. Tajik. The Echo Manual.
3rd Edition, Boston: Philadelphia: Lippincott Williams and Wilkins, (2007), p.70-72.
where:
RVOT SV = RVOT CSA* RVOT VTI = π/4 * RVOT diameter
LVOT SV = LVOT CSA * LVOT VTI = π/4 * LVOT diameter
Stroke Volume (SV) Doppler in ml
Oh, J.K., J.B. Seward, and A.J. Tajik. The Echo Manual. 3rd Edition, Boston: Philadelphia: Lippincott Williams
and Wilkins, (2007), p.69-71.
SV = (CSA * VTI)
where:
CSA = Cross Sectional Area of the orifice (LVOT area)
VTI = Velocity Time Integral of the LVOT

Cleaning and disinfecting

This section updates the instructions for cleaning and disinfecting the ultrasound system, transducers, and
accessories.
Use the FUJIFILM SonoSite recommendations when cleaning or disinfecting your ultrasound system,
transducer, and accessories. Use the cleaning recommendations in the peripheral manufacturer's instructions
when cleaning or disinfecting your peripherals.
Cleaning and disinfecting
* RVOT VTI
2
* LVOT VTI
2
7
Tabla de contenido
loading

Este manual también es adecuado para:

Sonosite sSonosite m-turbo

Tabla de contenido