Sony UP-D898MD Manual Del Usuario página 17

Ocultar thumbs Ver también para UP-D898MD:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 48
Guidance and manufacturer's declaration - electromagnetic immunity
The UP-D898MD is intended for use in the electromagnetic environment specified below. The customer or the user of the
UP-D898MD should assure that it is used in such as environment.
IEC 60601 test
Immunity test
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the UP-D898MD is used exceeds the applicable RF
compliance level above, the UP-D898MD should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the UP-D898MD.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Compliance level
level
3 Vrms
3 V/m
Electromagnetic environment-guidance
Portable and mobile RF communications equipment should
be used no closer to any part of the UP-D898MD, including
cables, than the recommended separation distance
calculated from the equation appliance to the frequency of
the transmitter.
Recommended separation distance
d = 1.2 √P
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2.5 GHz
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation distance
in meters (m).
Field strengths from fixed RF transmitters, as determined by
an electromagnetic site survey,
compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked
with following symbol:
should be less than the
a
b
5
GB
Tabla de contenido
loading

Tabla de contenido