Ottobock Genium 3B1-2 Instrucciones De Uso página 86

Ocultar thumbs Ver también para Genium 3B1-2:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 187
Beep signal
30x long
Error while charging the product
LED on
Status
power
LED on
supply
battery
charger
Status
LED
The LED
ring is lit
in weak
purple
86 | Ottobock
Vibration signal
1x long, 1x short
repeated every 3
seconds
Error
Country-specific plug adapter not
fully engaged on power supply
Non-functional socket
Defective power supply
No connection between battery
charger and power supply
Defective battery charger
Charging status
indicator (5 LEDs)
No LED is lit
Event
Severe error
e.g. failure of valve
drives
Possibly no switching
into safety mode (see
Page 80).
Resolution
Check whether the country-specif­
ic plug adapter is fully engaged on
the power supply.
Check socket with another electric
appliance.
The battery charger and power
supply must be inspected by an
authorised Ottobock Service
Centre.
Check whether the charging cable
plug is fully engaged on the battery
charger.
The battery charger and power
supply must be inspected by an
authorised Ottobock Service
Centre.
Error
Distance between bat­
tery charger and
receiver of the char­
ging unit on the pros­
thesis too great. If the
distance is more than
2 mm, the prosthesis
cannot be charged.
Required action
Restricted walking pos­
sible. The possible
change in flexion/exten­
sion resistance must be
noted.
Attempt to reset this error
by connecting/discon­
necting the battery char­
ger.
If the error persists, it is
no longer permissible to
use the product. The
product must be inspec­
ted immediately by a
prosthetist.
Resolution
Reduce distance
between battery char­
ger and receiver of the
charging unit.