Memory - Medisana MTV Instrucciones De Manejo

Esfigmomanómetro de brazo
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 47
5.1
Displaying
stored results
5.2
Deleting memory
6.1
Troubleshooting
Up to 60 measurements (blood pressure and pulse with time and date) may be
stored in each of the blood pressure monitor's two memories. The measured
values are stored automatically. Press the MEM button
measured values stored. The unit displays the memory set (U1 or U2) and the
number of memory locations occupied. Pressing the START button
switches to the other memory. The average measured values of the last three
measurements saved appear on the display if no button is pressed for a
duration of 5 seconds. An "A" (Average) appears instead of the memory
location number
. The data is announced in the language set if the
language function is switched on.
The values of the most recently saved measurement (memory location
number
e.g. "01") including the flashing blood pressure indicator
displayed after pressing the MEM button
during the measuring process, the arrhythmia symbol
voice function has been activated, the device speaks the memory location
number, together with the saved blood pressure and pulse values and the
blood pressure assessment as per the scale in the World Health Organisation
guidelines (see pg. 13).
Repeatedly pressing the MEM button
measured previously. In memory recall mode the unit switches off automa-
tically after 10 seconds if the last entry has been reached and the MEM button
has been pressed once more. Memory recall mode can be exited at any time
by pressing the START button
If all 60 memory locations in a memory are occupied, a new measured value
is saved whilst the oldest value is deleted.
If you are in memory recall mode, press and hold the MEM button
approximately 8 seconds. All measured data stored will be deleted automati-
cally after three beeps. Only noughts will be shown on the display instead of
the measured values.
The battery change symbol
• The batteries are too low or empty. Replace all four batteries with new 1.5 V
alkali AA batteries.
Erratic measurements being displayed
• Cuff incorrectly fitted or not tight enough. Fit cuff correctly and tighten.
• You have changed position or the cuff has moved during the measurement.
Keep still and do not move your arm or the cuff.
• You have an irregular heartbeat. Even if your heartbeat is only slightly
irregular, repeat the measurement. The unit is not suitable for people with
an extremely irregular heartbeat.
• You have spoken during the measurement, you are nervous or you are
excited. Do not speak and breathe deeply two or three times in order to
relax.
• You are not in the corrected seated or lying position. Please assume the
correct position when taking the measurement.
5 Memory / 6 Miscellaneous
. If an arrhythmia was detected
displays the respective values
, which will also switch the unit off.
appears in the display
GB
to call up the
are
also flashes. If the
for
17
Tabla de contenido
loading

Este manual también es adecuado para:

51120

Tabla de contenido