demineralised or distilled water to prevent calcium deposits.
method C: Sanitise the accessories (C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5-C6.1-C6.2-C6.3-
C6.4-C6.5-C6.6-C6.7.1-C6.7.2-C7) with a hot steam steriliser for baby-bottle (not the
microwave type). Perform the process faithfully following the instructions of the steriliser.
To ensure that the disinfection is effective, choose a steriliser with an operating cycle of at
least 6 minutes.
After having disinfected the accessories, shake them vigorously and place them on a
paper towel. Alternatively, dry them with a jet of hot air (for example, from a hair dryer).
At the end of each use store the device complete with accessories in a dry place away
from dust.
REPLACING THE FILTER
The device has a suction filter (A3) that should be replaced
when it is dirty or changes color. Do not wash or reuse the
same filter. Regular replacement of the filter is required to
help and ensure correct performance of the compressor.
Check the filter on a regular basis. Contact your local dealer
or authorized service center for spare filters.
Do not replace the filter during use.
Only use original accessories and spare parts by Flaem,
we disclaim any liability in the event of using non
original spare parts or accessories.
AVAILABLE SPARE PARTS
Description
- RF7 Dual Speed Nebulizer Set composed of: nebulizer and mouthpiece
- Mask Set composed of: adult mask, pediatric mask and elastic band
- 1 m connection tube
- Spare Air fi lter kit per unit (2 pcs)
- Nasal shower RHINO CLEAR®
ELECTROMAGNETIC COMPATIBILITY
This device was designed to satisfy the currently required requisites for electromagnetic
compatibility (EN 60601-1-2). Electro-medical devices require particular care during
installation and use relative to EMC requirements. Users are therefore requested to install
and/or use these devices following the manufacturer's specifications. Risk of potential
electromagnetic interference with other devices. Mobile or portable RF radio and
telecommunication devices (mobile phones or wireless connections) may interfere with
the operation of medical devices. For more information, visit the website www.flaemnuova.
it. The appliance may be susceptible to electromagnetic interference in the presence of
other devices used for specific diagnoses or treatments. Flaem reserves the right to make
technical and functional changes to the product without prior notice.