CLEANING SANITISATION DISINFECTION
Switch off the device before any cleaning procedure and unplug the power cable from
the socket.
DEVICE (A) AND TUBING EXTERIOR (B)
Use only a damp cloth with antibacterial soap (non-abrasive and with no solvents of any
sort).
ACCESSORIES
Disassemble the accessories as described below:
C1 - RF7 Dual Speed Nebulizer, Open the nebulizer by turning the top (C1.3)
counterclockwise, then remove the nozzle (C1.2) and the Speed selector (C1.4) from the
top (C1.3) and disassemble as shown in the "Assembly diagram" section C1.
C6 - Disassemble the NASAL SHOWER RHINO CLEAR® as shown in the "Assembly diagram"
section C6.
Open the nebuliser by turning the upper part (C1.3) anticlockwise, remove the nozzle
(C1.2) from the upper part (C1.3) by pressing as shown by the 2 arrows in the "Connection
diagram" in point C1.
Then proceed according to the following instructions.
SANITISATION
Before and after each use, sanitise the nebuliser cup and the accessories, choosing one of
the methods described below.
method A: Sanitise accessories (C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5-C6.1-C6.2-C6.3-C6.4-
C6.5-C6.6-C6.7.1-C6.7.2-C7) under potable hot water (approximately 40°C) with a gentle,
non abrasive dish detergent.
method B: Sanitise accessories (C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5-C6.1-C6.2-C6.3-C6.4-
C6.5-C6.6-C6.7.1-C6.7.2-C7) in the dishwasher with a hot cycle.
method C: Sanitise accessories (C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5-C6.1-C6.2-C6.3-C6.4-
C6.5-C6.6-C6.7.1-C6.7.2-C7) by immersing them in a solution of 50% water and 50% white
vinegar, then rinse thoroughly under potable hot water (approximately 40 °C).
If you want to also perform the cleaning for DISINFECTION, jump to the DISINFECTION
paragraph.
After having sanitised the accessories, shake them vigorously and place them on a paper
towel. Alternatively, dry them with a jet of hot air (for example, a hair dryer).
DISINFECTION
After sanitising the nebuliser cup and the accessories, disinfect them choosing one of the
methods described below.
method A: Accessories (C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5-C6.1-C6.2-C6.3-C6.4-C6.5-
C6.6-C6.7.1-C6.7.2-C7) can be disinfected.
The disinfectant must be an electrolytic chloroxidizer (active principle: sodium
hypochlorite) specific for disinfecting, which is available in any pharmacy.
Implementation:
- Fill a container big enough to hold all of the parts to disinfect with a solution of potable
water and disinfectant, according to the proportions indicated on the packaging of the
disinfectant.
- Completely immerse each part in the solution, taking care to avoid the formation of air
bubbles on the parts. Leave the parts immersed for the amount of time indicated on the
packaging of the disinfectant associated with the concentration chosen for the solution.
- Remove the disinfected parts and rinse abundantly with warm potable water.
- Dispose of the solution following the instructions provided by the disinfectant
manufacturer.
method B: Sanitise the accessories (C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5-C6.1-C6.2-
C6.3-C6.4-C6.5-C6.6-C6.7.1-C6.7.2-C7) by boiling them in water for 10 minutes; use
14