packaging of the disinfectant associated with the concentration chosen for the solution.
- Remove the disinfected parts and rinse abundantly with warm potable water.
on
- Dispose of the solution following the instructions provided by the disinfectant
or
manufacturer.
he
method B: Sanitise the accessories C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5 by boiling them
in water for 10 minutes; use demineralised or distilled water to prevent calcium deposits.
u-
method C: Sanitise the accessories C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5 with a hot steam
r
steriliser for baby-bottle (not the microwave type). Perform the process faithfully following
-
the instructions of the steriliser. To ensure that the disinfection is effective, choose a
d
steriliser with an operating cycle of at least 6 minutes.
-
After having disinfected the accessories, shake them vigorously and place them on a
-
paper towel. Alternatively, dry them with a jet of hot air (for example, from a hair dryer).
At the end of each use store the device complete with accessories in a dry place away
from dust.
he

REPLACING THE FILTER

The device has a suction filter (A3) that should be replaced when
ny
it is dirty or changes color. Do not wash or reuse the same filter.
Regular replacement of the filter is required to help and ensure
correct performance of the compressor. Check the filter on a
zle
regular basis. Contact your local dealer or authorized service
on
center for spare filters.
Do not replace the filter during use.
Only use original accessories and spare parts by Flaem, we disclaim any liability in
the event of using non original spare parts or accessories.
of
ot
th
in
ot
ON
er
he
um
ble
he
air
he
13