Error
Cause
description
Readings not transmitted
Err 4
correctly.
Power pack or battery
voltage too low.
Err 5
Battery contacts are
corroded.
Air congestion
Err 6
Blood pressure cuff
incorrectly connected.
Leak in the blood
pressure cuff or air hose
Remedy
Wait until you take another measurement. The data are then
transmitted together with the new readings.
If this error occurs repeatedly, send the blood pressure monitor to
your healthcare provider or directly to the manufacturer for
checking.
➔
Change the batteries (
Chapter 5.2).
Clean the battery contacts with a cotton cloth and a little alcohol.
Check the blood pressure cuff for air congestion or a kink in the air
hose. If there is a kink in the air hose, unkink it, otherwise send the
blood pressure monitor to your healthcare provider or the
manufacturer.
Connect the blood pressure cuff to the blood pressure monitor
➔
(
Chapter 6.2).
Replace the blood pressure cuff with the air hose.
Error messages
EN
37