Cleaning and disinfection
WARNING
▪
When putting the blood pressure monitor
on,
there
must
disinfectants on the blood pressure cuff!
▪
There are patients who have intolerances
(e.g. allergies) to disinfectants or its
components!
ATTENTION
▪
Do not immerse the cuff with balloon and
blood pressure monitor in disinfectant,
water or other liquids!
▪
If liquid nevertheless penetrates the device,
switch it off immediately and send it away
to be checked by the manufacturer or
healthcare provider!
▪
Do not open the housing of the Tel-O-
®
Graph
GSM, otherwise any warranty
becomes void.
no
longer
be
any
Note
Always observe the manufacturer's instructions
on the disinfection and cleaning of these
products.
9.1
Cleaning
ATTENTION
▪
To clean, use lukewarm water up to 30 ° C
at most, to which you can add a mild
detergent
if
necessary.
abrasive or solvent-based detergents as
these can damage the surface of the blood
pressure monitor and the cuff.)
▪
Do not use fabric softeners or other
additives (e.g. hygiene rinses, textile
deodorants). These agents can leave
behind residues and damage the material.
▪
The cuff can be washed in the washing
machine up to 30° C using a mild detergent
without spin-dry.
▪
The cuff is not suitable for drying in a dryer.
▪
The Velcro strip must always be closed
before washing.