7.
Sagittal saw attachment
GB660R
Abb. 6

7.1 Intended use

The sagittal saw attachment is used for intraoperative
dissection of bone and cartilage tissue.

7.2 Operating principle

The sagittal saw attachment is powered by a motor
fitted with the Aesculap Acculan® 3Ti coupling.
The sagittal saw attachment converts the motor
rotation into oscillation of the saw blade. The bone is
dissected by the oscillating movement of the saw
blade.
No key or wrench is required for inserting the saw
blades. The sagittal saw attachment can be coupled to
the drive unit in three different positions separated by
steps of 120°.

7.3 Safe operation

Risk of injury caused by inadvertent
loosening of the saw blade during
operation!
Do not press the button for saw
WARNING
blade
operation.
Check for secure fixation of the
tool after any tool change.
release/fixation
during
Connect the sagittal saw attachment, siehe Cou-
pling and uncoupling of attachments on the drill
and reamer motor unit.

Coupling the saw blade

Insert the saw blade with its coupling end in the
slot of saw blade coupling 7.1 as far as it will go.
Ensure that the locking pin engages in the saw
blade fenestration, and that the lateral stops of the
saw blade rest on the coupling.
If necessary press button for saw blade release/fix-
ation 7.2.
Check that the tool is seated properly: Pull at the
saw blade.

Uncoupling the saw blade

Fully
press
button for saw
fixation 7.2.
Extract the saw blade from saw blade coupling 7.1.

7.4 Technical specifications

Dimensions L x W x H, in
mm
Weight, in grams
Gear ratio
Stroke frequency in 1/min
blade
release/
GB660R
89 x 36 x 44
approx. 270
1.5:1
max. 17 000
11