EN-2
2.2 Scale Bases
The T82XWT indicator supports analog scale bases and provides 5 volts of excitation to drive analog load cells. It can be
connected to 4 platforms and up to 16 350Ω load cells can be powered by the indicator.
A six wire load cell connection is provided with sense lines to help maintain accuracy as the load cell cable resistance
changes with temperature variations.
2.3 Opening the Enclosure
The front panel of the harsh enclosure T82XWT indicator is locked in place by six screws attached to the body of the
enclosure. To access the indicator's PCB to install options, connect internal wiring and set switches, separate the front panel
from the enclosure by unscrewing the six screws:
2.5.1 Main Board Wiring Connections
Once the T82XWT indicator harsh enclosure is open, connections can be made to the indicator strips on the main board, as
shown below.
Load cell
connectors
Opening the Harsh Enclosure
Serial Ports
T82XWT Main Board Connections
7000 SERIES INDICATO
Security
switch
cover