Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The Connected OR Cart, model 240–099–155 and 240-099-156, is intended for use in the electromagnetic environment speci-
fied below. The customer or the user of the Connected OR Cart should assure that it is used in such an environment.
Immunity Test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interrup-
tions and voltage variations on
power supply input lines
IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
Data extracted from IEC 60601-1-2:2014, Table 4 and Table 5
16 EN
IEC 60601 Test Level
± 8 kV contact
± 15 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
0% Ut 0.5 cycle
0% Ut 1 cycle
70% Ut 25 cycles
0% Ut 5 Sec
30 A/m
Compliance Level
± 8 kV contact
± 15 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
0% Ut 0.5 cycle
0% Ut 1 cycle
70% Ut 25 cycles
0% Ut 5 Sec
30 A/m
Electromagnetic Environment
— Guidance
Floors should be wood, con-
crete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%.
If disruption of video occurs,
the device may require restart.
Mains power quality should be
that of a typical commercial or
hospital environment.
If disruption of video occurs,
the device may require restart.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial
or hospital environment. If the
user of the Connected OR
Cart, model 240–099–155,
requires continued operation
during power mains interrup-
tions, it is recommended that
the Connected OR Cart be
powered from an uninterrupt-
ible power supply.
Power frequency magnetic
fields should be at levels char-
acteristic of a typical location
in a typical commercial or
hospital environment.
P44082 Rev. D