Guidance and- manufacturer's declaration.
The device is intended for use In. the electromagnetic
environment specified below. The customer or the user
should assure that It is used in such an environment.
Immunity test IEC 60501
test level
Conducted
3 Vrms
RF lEC
150 kHz
61000-4-6
to 80 MHz
Radiated
3 V/m80
RF lEC
MHz to
61000-4-3
2.5 GHz
All manuals and user guides at all-guides.com
Electromagnetic immunity
Compliance
level
3 Vrms
3 V/m
37
Electromagnetic
environment — guidance
Portable and mobile RF
Communications
equipment should be
used no closer to any
part of the device,
including cables, than the
recommended separation
distance calculated from
the equation applicable
to the frequency of the
transmitter.
Recommended
separation distance
80 MHz
a 800 MHz
800 MHz
a 2.5 MHz