intensity TWIN STIM III Manual De Instrucciones página 16

Ocultar thumbs Ver también para TWIN STIM III:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Technical specifications for Transcutaneous Electrical Nerve
Stimulator (TENS) mode
Waveform
Pulse amplitude Adjustable, 0—lO5mA at 1000 ohm Load each
Pulse Width
Pulse Rate
Burst (B)
Normal (N)
Pulse Width
Modulation (M)
Pulse Rate
Modulation (M1)
All manuals and user guides at all-guides.com
Mono-phase square pulse wave
channel, imA/step
Adjustable, from 50 to 300ps, lOps/step.
Adjustable, from I to 150 Hz, 1 Hz/step
Burst rate: Adjustable, 0.5 – 5Hz
Pulse width adjustable, 50 – 300ps
Frequency fixed = 10Hz
The pulse rate and pulse width are adjustable.
It generates continuous stimulation based on
the setting value.
The pulse width is automatically varied in a cycle
time. The pulse width is decreased from its original
setting to 60% in setting cycle time, and then
increased from 60% to its original setting in nest
setting cycle time. In this program, pulse rate
(1 – 150Hz), pulse width (50 – 300ps)
and cycle time (5 – 30 sec) are fully adjustable.
The pulse rate is automatically varied in a cycle
time. The pulse rate is decreased from its original
setting to 60% in setting cycle time, and then
increasing from 60% to its original setting in nest
setting cycle time. In this program, pulse rate
(1 – 150Hz), pulse width (50 – 300ps) and cycle
time (5 – 30 sec) are fully adjustable.