Adjust Channel Intensity; Safety Lock Feature - intensity TWIN STIM III Manual De Instrucciones

Ocultar thumbs Ver también para TWIN STIM III:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Caution:
Contract time does not include the ramp up and ramp down
time; ON time = Ramp up + Contract time + Ramp down.
8) Set Relaxation (OFF) time
The Off Time controls the time of relaxation. The relaxation time can
be adjusted. Press [S] button cycle to enter this menu, and then
press the [▲] and [▼] button to adjusting the setting. Both channels'
stimulation is cycled on and off by the contraction and relaxation
settings. The range is adjustable from 0 – 60 seconds.
In Alternate program, the OFF Time should be equal or more than
the ON Time. (OFF TIME ≥ ON TIME)

4~8 Adjust Channel Intensity

Press the intensity control button ([▲] and [▼]) to control the
intensity output. Slowly press the intensity button control until you
reach the setting recommended by your physician or therapist.
Repeat for the other channel, if both channels are to be used.
Caution:
1)
If the stimulation levels are uncomfortable or become
uncomfortable, reduce the stimulation intensity to a
comfortable level and contact your medical practitioner if
problems persist.
2)
If the electrodes are not placed firmly on the skin or the
device is not connected to the electrodes, and the
stimulator's output intensity surpasses l2mA, the amplitude
level will be reset to 0 mA automatically.

4.9. Safety Lock Feature

The Safety Lock Feature automatically activates after there is no
operation in the panel for 30 seconds by locking out the ability to
press the buttons. This is a safety feature to prevent accidental
changes to you settings and to prevent accidental increases to the
intensity levels. You can press either one of the [▲] button to unlock
the device.