Electrical fast
± 2 kV for power
transient/burst
supply lines
IEC 61000-4-4
±1 kV for input/
output lines
Surge
± 1 kV
IEC
differential mode
61000-4-5
± 2 kV common
mode
Voltage
< 5% UT (>
dips, short
95% dip in UT)
interruptions
for 0.5 cycle
and voltage
variations on
40% UT (60%
power supply
dip in UT) for 5
input lines IEC
cycles
61000-4-11
70% UT (30%
dip in UT) for 25
cycles
< 5% UT (>
95% dip in UT)
for 5 seconds
Power
3 A/m
frequency
(50/60 Hz)
magnetic
field IEC
61000-4-8
NOTE UT is the a.c mains voltage prior to application of the test level.
All manuals and user guides at all-guides.com
Not
Applicable
Not
Applicable
Not
Applicable
3 A/m
36
Mainspower quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment
Mains power quality
should be that of a
typical commercial or
hospital environment
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment