S5 Apex Locator is intended for use in professional healthcare facility or home healthcare
electromagnetic environment specified below; The customer or the user of S5 should assure that it is
used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
S5 Apex Locator is intended for use in the electromagnetic environment specified below;
The customer or the user of S5 Apex Locator should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transients/bursts
IEC 61000-4-4
Surges
IEC 61000-4-5
Guidance and manufacturer's declaration:
electromagnetic emissions -
Compliance
Group 1
Class B
Complies
Complies
Guidance and manufacturer's declaration:
electromagnetic immunity - S5 Apex Locator
IEC 60601-1-2
Test level
±8 kV contact
±15 kV air
±2 kV for power
supply lines
±1 kV for input/output
lines
±1 kV
Line-to-line
±2 kV
Line-to-ground
S5 Apex Locator
Electromagnetic environment - guidance
S5 Apex Locator uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
S5 Apex Locator is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low voltage
power supply network that supplies buildings used for
domestic purposes.
Compliance level
±8 kV contact
±15 kV air
Not Applicable
Not Applicable
22
Electromagnetic
environment - guidance
Floors should be wood,
concrete or Ceramic tile. If
floors are covered with
synthetic material, the relative
humidity Should be at least
30%.
Not Applicable
Not Applicable