Erbe 20132-200 Manual Del Usuario página 33

Vio apc-mangos
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 58
OBJ_DOKU-188704-009.fm Seite 33 Donnerstag, 25. März 2021 11:48 11
The detergent and disinfectant must be suitable for medical de-
vices made of plastic and metal and have a pH value between 5.5
and 12.3.
Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter "Overview of validated
procedures".
1. Prepare a cleaning bath and a separate disinfectant bath.
2. Immerse the product completely in a cleaning bath, ensuring
that the product does not come into contact with other com-
ponents in the bath.
3. Thoroughly clean the surfaces with a soft brush/soft dispos-
able cloth.
4. Completely fill a disposable syringe with cleaning solution
and rinse the lumina (at least 5 times).
5. Rinse the surfaces with sterile, deionized water (at least
1 minute).
6. Completely fill a disposable syringe with sterile, deionized
water and rinse the lumina (at least 5 times).
7. Repeat the preceding cleaning steps if the product or used
rinse water still contains visible contamination.
8. Immerse the product completely in a disinfectant bath, en-
suring that the product does not come into contact with oth-
er components in the bath.
9. Completely fill a disposable syringe with disinfecting solution
and rinse the lumina (at least 5 times).
10. Rinse the surfaces with sterile, deionized water (at least
1 minute).
11. Completely fill a disposable syringe with sterile, deionized
water and rinse the lumina (at least 5 times).
12. Remove the irrigation adapter.
13. Dry the product until there are no more fluid residues visible
(e.g. with filtered compressed air).
33
Tabla de contenido
loading

Este manual también es adecuado para:

20132-201

Tabla de contenido