M/L-10 Clip Applier Sterilization Tray (Catalog # 3706) - Instructions For Use
INDICATIONS FOR USE
Microline Surgical M/L-10 Clip Applier Sterilization Tray (Catalog# 3706) is intended to protect the M/L-10 Clip Applier handpiece (Catalog #1002)
and to facilitate the sterilization process by allowing steam penetration and air removal. When used in conjunction with 2 layers of FDA-cleared
1-ply polypropylene sterilization wrap, sterility of the enclosed sterilization tray (Catalog# 3706) is maintained until used.
The Microline Surgical M/L-10 Clip Applier Sterilization Tray (Catalog# 3706) is used to secure the M/L-10 Clip Applier handpiece (Catalog
#1002) which measures at 17.13" length, 4.94" height, and 0.90" width in dimensional measurements, as the intended load during the transport,
sterilization and storage between its intended uses.
The M/L-10 Clip Applier Sterilization Tray (Catalog# 3706) has been validated with the M/L-10 Clip Applier handpiece (Catalog #1002) wrapped
in 2 layers of FDA-cleared 1-ply polypropylene wrap weighing a calculated total weight of 1663.10 grams. The M/L-10 Clip Applier Sterilization
Tray (Catalog# 3706) has been validated with the following Steam Sterilization Parameters: Pre-vacuum, Time: 4 minutes at Temperature of
132 °C. The Gravity Cycle parameters include: Time: 15 minutes and Temperature of 132 °C.
CONTRAINDICATIONS
The M/L-10 Clip Applier Sterilization Tray (3706) is not intended for use except as indicated.
COMPONENTS OF A M/L-10 CLIP APPLIER STERILIZATION TRAY
The M/L-10 Clip Applier Sterilization Tray (3706) consists of:
1. One perforated stainless steel Lid.
2. One perforated stainless steel Base.
3. Silicone brackets.
4. Alluminum brackets.
PRE-CHECK INSTRUCTIONS
1. Carefully inspect the packaging to ensure that there are no signs of damage. If damage is visible, do not use the sterilization tray (3706).
2. Ensure that the sterilization tray (3706) received is utilized for the M/L-10 Clip Applier device only as specified by the supplied IFU.
DEVICE DESCRIPTION
The Microline Surgical M/L-10 Clip Applier sterilization tray (Catalog #3706) is used to secure the M/L-10 Clip Applier device as intended load
during the transport, sterilization and storage between intended use. The sterilization tray (3706) consists of a stainless steel case, a stainless
steel cover and silicone and aluminum brackets. The cover is secured to the case by means of assembled hardware and a locking tab, designed
as a part of the lid cover. The brackets are designed to secure the M/L-10 Clip Applier for sterilization. The tray cover and case are designed
using stainless steel, which can be reused with steam sterilization methods. The brackets are constructed from aluminum, stainless steel, and
biomedical grade silicone. The case and cover have an evenly distributed hole pattern in relation to its size. The M/L-10 Clip Applier sterilization
tray (3706) has been validated weighing 1663.10 grams for cleaning and sterilization cycles. The validated prevacuum cycle parameters are
as follows: Time: 4 minutes, and temperature:132 °C.The gravity cycle parameters are as follows: Time: 15 minutes, and temperature: 132 °C.
These cleaning and sterilization cycles are validated with the sterilization tray (3706) wrapped in 2 layers of FDA-cleared 1-ply polypropylene
wrap. The M/L-10 Clip Applier sterilization tray (Catalog #3706) is intended for over-the-counter use only.
PREPROCESSING INSTRUCTIONS
Prior to use, follow the guidelines below for applicable cleaning and sterilization methods.
MANUAL CLEANING (minimum required steps)
1. Rinse with running tap water to remove gross soil.
2. Prepare enzymatic detergent solution in accordance to the manufacturer's recommendations.
3. Immerse in enzymatic solution; allow to soak for a minimum of five (5) minutes.
4. Clean the device of any visible signs of contamination using a soft bristled brush.
5. Rinse with lukewarm running tap water.
6. Rinse with deionized water. Repeat 3 times.
7. Dry completely with a clean, soft, lint-free cloth.
8. Visually inspect to ensure all soil is removed.
AUTOMATED CLEANING
1. Rinse with running tap water to remove gross soil.
2. Load device(s) into automated washer. Load devices(s) in the open position.
3. Automated wash cycle parameters as follows:
Instrument Cycle: High
Phase
Time (Minutes)
Pre-Wash
2:00
Wash 1
2:00
Rinse 1
00:15
PURW Rinse
1:00
Dry
6:00
4. Visually inspect to ensure all soil is removed.
5. Package appropriately for sterilization.
STERILIZATION
Sterilization Method
Temperature
Pre-vacuum
132 °C
Gravity
132 °C
INSTRUCTIONS FOR USE
1. Clean and disinfect the used M/L-10 Clip Applier device (1002) as per the M/L-10 Clip Applier IFU (IFU0030), then secure M/L-10 Clip Applier
device in the silicone and aluminum brackets of the sterilization tray (3706).
2. Cover the base with the lid and secure the locking tab.
3. Wrap the sterilization tray with 2 layers of FDA-cleared 1-ply polypropylene wrap.
4. Sterilize the loaded wrapped tray in accordance with the M/L-10 Clip Applier device manufacturer's guidelines for steam sterilization methods.
5. After steam sterilization, move the wrapped sterilization tray to a sterile storage area for storage or transport to the point of use location.
6. Return the sterilization tray (3706) to the sterile processing department after use.
7. The sterilization tray (3706) is reusable and should be cleaned and decontaminated prior to reuse.
STORAGE CONDITIONS
After sterilization, product should be stored in a clean, dry place.
ENVIRONMENTAL PROTECTION
To reduce contamination risks, the M/L-10 Clip Applier Sterilization Tray (3706) shall be disposed of in compliance with all applicable local, state
and federal laws and regulations.
WARRANTY
Microline Surgical warrants that its instruments are free from any defects in both material and workmanship. Microline Surgical shall not be held
liable for any incidental or consequential damage of any kind. Work performed on an instrument will void this warranty. Gross abuse or neglect
of a Microline Surgical instrument will void this warranty.
M/L-10 Clip Applier Sterilization Tray
Instructions For Use
Microline Surgical, Inc.
50 Dunham Road, Suite 1500
Beverly, MA 01915 U.S.A.
Tel: +1(978) 922-9810 | Fax: +1(978) 922-9209
Web: www.microlinesurgical.com
09-39-00238-00-ART Rev A 2/2016
Temperature
Detergent
Cold Tap Water
N/A
65.5 °C
Alkaline Detergent
Hot Tap Water
N/A
90 °C
N/A
98.8 °C
N/A
Sterilization Parameters
Time
Sterilization Wrap
Wrapped in two layers of 1-ply
4 minutes
polypropylene wrap of appropri-
ate basis weight to accomodate
the weight of the sterilization tray.
15 minutes
ENGLISH
Tabla de contenido
loading

Resumen de contenidos para Microline Surgical M/L-10

  • Página 1 The cover is secured to the case by means of assembled hardware and a locking tab, designed as a part of the lid cover. The brackets are designed to secure the M/L-10 Clip Applier for sterilization. The tray cover and case are designed using stainless steel, which can be reused with steam sterilization methods.
  • Página 2: Symbol Definition

    REUSE LIFE Due to the M/L-10 Clip Applier sterilization tray (3706) design, the raw materials used and the intended purpose, it is not possible to determine a precise limit with regard to the maximum possible number of reprocessing cycles. The serviceable life of the instruments is determined by their function as well as by careful handling.
  • Página 3: Contre-Indications

    Le plateau de stérilisation pour pince à clips M/L-10 (n° réf. 3706) de Microline Surgical est utilisé pour fixer la pièce à main de la pince à clips M/L-10 (n° réf. 1002), qui mesure 43,51 cm (17,13 po) de long, 12,55 cm (4,94 po) de haut et 2,29 cm (0,90 po) de large, et supporter son poids pendant le transport, la stérilisation et le stockage, entre ses utilisations prévues.
  • Página 4: Durée De Réutilisation

    DURÉE DE RÉUTILISATION En raison de la conception du plateau de stérilisation pour pince à clips M/L-10 (n° réf. 3706), des matières premières utilisées et de l’utilisation prévue, il n’est pas possible de déterminer un nombre maximal précis de cycles de retraitement. La vie utile des instruments est déterminée par leur fonction, ainsi que par le soin avec lequel ils sont manipulés.
  • Página 5 Il vassoio di sterilizzazione per l'applicatore di clip M/L-10 Microline Surgical (n. di catalogo 3706) viene utilizzato per proteggere il manipolo dell'applicatore di clip M/L-10 (n. di catalogo 1002) che ha una lunghezza di 43,51 cm (17,13”), un'altezza di 12,55 cm (4,94”) e una larghezza di 2,29 cm (0,90”), come carico previsto durante il trasporto, la sterilizzazione e la conservazione tra un utilizzo e l'altro.
  • Página 6: Definizioni Dei Simboli

    DEFINIZIONI DEI SIMBOLI Qui di seguito sono elencati i simboli grafici usati nelle etichette del vassoio di sterilizzazione per l'applicatore di clip M/L-10 con le rispettive descrizioni. Il prodotto non contiene lattice di gomma naturale.
  • Página 7 DEUTSCH INDIKATIONEN Der Sterilisationsbehälter für den M/L-10 Clipapplikator von Microline Surgical (Bestell-Nr. 3706) ist für den Schutz des M/L-10 Clipapplikator- Handstücks (Bestell-Nr. 1002) vorgesehen sowie zur Erleichterung des Sterilisationsverfahrens durch den Einlass von Dampf und den Auslass von Luft. Bei Verwendung in Verbindung mit 2 Lagen FDA-zugelassener 1-schichtiger Polypropylensterilisationsfolie bleibt der umschlossene Sterilisationsbehälter (Bestell-Nr.
  • Página 8 Sterilisationsbehälters (3706) von der individuellen intraoperativen Anwendung und den spezifischen Krankenhausvorschriften zur Wiederaufbereitung ab. SYMBOLDEFINITIONEN Im Folgenden sind die grafischen Symbole aufgeführt und beschrieben, die zur Kennzeichnung des Sterilisationsbehälters für den M/L-10 Clipapplikator verwendet werden: Nicht mit Naturlatex hergestellt.
  • Página 9: Esterilización

    La bandeja de esterilización para aplicador de clips M/L-10 de Microline Surgical (n.º de catálogo 3706) sirve para proteger la pieza de mano del aplicador de clips M/L-10 (n.º de catálogo 1002) —que mide 43,51 cm (17,13”) de largo, 12,55 cm (4,94”) de alto y 2.29 cm (0,90”) de ancho—...
  • Página 10: Período De Reutilización

    PERÍODO DE REUTILIZACIÓN Dado el diseño de la bandeja de esterilización para aplicador de clips M/L-10 (3706), la materia prima utilizada y el fin previsto, no es posible determinar con exactitud el número máximo posible de ciclos de reprocesamiento. La vida útil de los instrumentos depende de su función y de su manipulación cuidadosa.
  • Página 11 NEDERLANDS GEBRUIKSINDICATIES De Microline Surgical sterilisatietray voor M/L-10 clip-applicator (catalogusnr. 3706) is geïndiceerd ter bescherming van het handstuk van de M/L-10 clip-applicator (catalogusnr. 1002) en bevordert het sterilisatieproces door stoompenetratie en luchtverwijdering toe te laten. Wanneer de bijgeleverde sterilisatietray (catalogusnr. 3706) wordt gebruikt in combinatie met twee lagen sterilisatiewikkel gemaakt van door de (Amerikaanse) FDA goedgekeurd eenlagig polypropyleen, blijft deze steriel tot het moment van gebruik.
  • Página 12: Definitie Van De Symbolen

    LEVENSDUUR BIJ HERGEBRUIK Vanwege het productontwerp, de gebruikte grondstoffen en het beoogde gebruik van de sterilisatietray voor M/L-10 clip-applicator (3706) kan de limiet met betrekking tot het maximumaantal herverwerkingscycli niet precies worden vastgesteld. De gebruiksduur van de instrumenten hangt af van hun functie en de mate waarin er zorgvuldig mee is omgegaan. De sterilisatietray (3706) is gevalideerd voor honderd (100) herverwerkingscycli.
  • Página 13: Contraindicações

    O tabuleiro de esterilização (3706) é composto por uma caixa de aço inoxidável, uma tampa de aço inoxidável e suportes de silicone e alumínio. A tampa é fixa à caixa através de ferragens e de um fecho nela integrado. Os suportes foram concebidos para fixar o Aplicador de clipes M/L-10 durante a esterilização.
  • Página 14: Definição Dos Símbolos

    DURAÇÃO DA REUTILIZAÇÃO Devido ao seu desenho, matérias primas utilizadas e finalidade a que se destina o Tabuleiro de esterilização para aplicador de clipes M/L-10 (3706), não é possível determinar um limite preciso em relação ao número máximo possível de ciclos de reprocessamento. A vida útil dos instrumentos depende da sua função e do cuidado no seu manuseamento.
  • Página 15 Microline Surgical M/L-10 Klips Uygulayıcı Sterilizasyon Tepsisinin (Katalog No: 3706), boyutları 43,51 cm (17,13 inç) uzunluk, 12,55 cm (4,94 inç) yükseklik ve 2,29 cm (0,90 inç) genişlik olan M/L-10 Klips Uygulayıcı el kısmını (Katalog No: 1002) amaçlanan kullanımları arasındaki taşıma, sterilizasyon ve saklama sırasında amaçlanan yük olarak sabitlemesi amaçlanmıştır.
  • Página 16 YENİDEN KULLANIM ÖMRÜ M/L-10 Klips Uygulayıcı Sterilizasyon Tepsisi (3706) tasarımı, kullanılan ham maddeler ve belirlenen amaç nedeniyle maksimum tekrar işleme döngüsü sayısı açısından kesin bir sınır belirlemek mümkün değildir. Aletlerin hizmet ömrü, işlevleriyle ve ayrıca dikkatli muameleyle belirlenir. Sterilizasyon tepsisi (3706) 100 kullanıma kadar tekrar işleme döngüsü doğrulamasından geçmiştir. Sterilizasyon tepsisinin (3706) her amaçlanan kullanımı...
  • Página 17 .‫النقل والتعقيم والتخزين بين عمليات استخدام األداة المخصصة لها. تتكون صينية التعقيم )6073( من علبة من الصلب المقاوم للصدأ وغطاء من الصلب المقاوم للصدأ و م ُ ثبتات من السليكون واأللومنيوم‬ ‫ إلجراء التعقيم. تم تصميم غطاء الصينية وعلبتها‬M/L-10 ‫يكون الغطاء مثب ت ً ا بالعلبة عن طريق أجزاء صلبة مجمعة ولسان إغالق مصمم كجزء من الغطاء. تم تصميم ال م ُ ثبتات لتثبيت أداة تثبيت المشابك‬...
  • Página 18 ‫تعريفات الرموز‬ M/L-10 ‫فيما يلي الرموز الرسومية المستخدمة، وأوصافها، الموجودة على ملصق صينية تعقيم أداة تثبيت المشابك‬ .‫غير مصنوع من مادة المطاط الطبيعي المرن‬ ‫رقم الكتالوج‬ .‫ال ي ُستخدم في حالة تلف العبوة‬ ‫رقم التشغيلة‬ ‫راجع تعليمات االستخدام‬ ‫الرقم المسلسل‬...

Tabla de contenido