  • ESPAÑOL, página 93

6.1 Func-Stor

Manual recording of measurement values
In this mode, up to 1,000 records can be saved. To do this, select the item Func = Stor in
the configuration menu <SEt LoGG>. An indicator arrow ( ) now appears on the display
above logg. From now on, measurement data can be stored in operating mode as follows:
/
i
If the logger memory is full, the following message appears on the
display: Data are retrieved and deleted via the data archive.

6.2 Func-CYCL

Automatic recording of measurement values
In this mode, up to 10,000 records can be saved. To do this, select the item Func = CYCL in
the Configuration menu <SEt LoGG>. An indicator arrow ( ) now appears on the display
above logg. Automatic data recording at intervals of the set cycle time can be started by
button press. To do this, proceed as follows:
Data are deleted via the data archive.
60
Briefly pressing the <Store> button stores a record in operating mode
Select a measurement ID from "L ID" (0 ... 9999). This function allows
you to assign the measured value to a sample or measuring location.
Confirm the entry
Start measurement value recording:
A long press on the <Store> button in operating mode causes the
<Logg Run> prompt to appear. Confirm with
The indicator arrow above logg starts to flash > <
Stop measurement value recording:
A long press on the <Store> button during data recording causes
the <Logg Stop> prompt to appear. Confirm with
.
.