6.3
Required aids
Reprocessing step Items Needed
Precleaning
Soft plastic brush/soft cloth
Manual cleaning/
Soft plastic brush/soft disposable cloth
disinfection
(low particle)
Compressed air gun (2 bar maximum)
6.4
Precleaning
Use water for precleaning, a non-fixing disinfectant if necessary.
For heavy soiling Erbe Elektromedizin recommends an approx.
3-percent hydrogen peroxide bath.
When using a soft plastic brush, ensure that the brush is moved
only in parallel with the coagulation surface of the BiClamp or
from the backs of the jaws in the direction towards the coagula-
tion surface. On no account brush from the coagulation surface
in the direction of the backs of the jaws. See also the following
illustration:
1. If required, place the product in an approx. 3-percent hydro-
gen peroxide bath.
24