Precleaning The Product In The Procedure Room; Cleaning, Disinfection, Sterilization; Safety Instructions; Reprocessing Limitation - Erbe BiClamp 20195-200 Manual De Instrucciones

Ocultar thumbs Ver también para BiClamp 20195-200:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 46
OBJ_DOKU-193005-006.fm Seite 23 Mittwoch, 31. Oktober 2018 12:40 12
5.7

Precleaning the product in the procedure room

1. Use a soft cloth (such as a compress) to remove any obvious
contamination immediately after use and/or place the prod-
uct in deionized water.
6

Cleaning, disinfection, sterilization

6.1

Safety instructions

May be cleaned in an ultrasonic bath.
On no account use sharp or abrasive objects for cleaning purpos-
es.
Disinfectants must be rinsed off thoroughly after use.
Maximum air pressure during drying: 2 bar.
Maximum water pressure during rinsing: 2 bar.
During machine cleaning and disinfection, do not exceed 95 °C.
Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter "Overview of validated
procedures".
During sterilization, do not exceed 138 °C.
Do not sterilize in hot air.
Do not sterilize with gamma radiation.
6.2

Reprocessing limitation

Frequent reprocessing can damage this product. Observe the
safety instructions for checking the product. With visible damage
or functional impairment, the product may no longer be used.
23
Tabla de contenido
loading

Tabla de contenido