Table 4
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Recommended separation distances between
Portable and mobile RF communications equipment and the [KD-595]
The [KD-595] is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the [KD-595] can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the [KD-595] as recommended below,
according to the maximum output power of the communications equipment
Rated maximum output
power of transmitter
W
150 kHz to 80 MHz
80 MHz to 800 MHz
d = 1.2 √P
d = 1.2 √P
0.01
0.12
d = 1.2 √P
0.1
0.38
1
1.2
10
3.8
100
12
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can
be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating
of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
114
Visiomed - KD-595 - Manuel d'utilisation - 072017
Separation distance according to frequency of transmitter
m
800 MHz to 2,5 GHz
d = 2.3 √P
0.23
0.38
0.73
1.2
2.3
3.8
7.3
12
23
Visiomed - KD-595 - Manuel d'utilisation - 072017