Specifications, Blood Pressure Measurement
Measurement:
Range:
Accuracy:
Conditions for
Use:
Power:
Calibration:
Safety Systems: Independent hardware over-pressure circuit and redundant
Dimensions:
Auscultatory, using R-wave gating and k-sound analysis, during
all static and active phases of stress testing. Systolic pressures
correlate to a K-1 Korotkoff sound. Diastolic pressures correlate
to K-5 Korotkoff sound.
The device is designed to function in the presence of a normal
ECG sinus rhythm. There are some physical conditions (i.e.
Bundle Branch Block, Arrhythmias, Arial Fibrillation, Ventricular
Fibrillation, Pacemakers, etc) that may limit the ability of Tango+
to obtain an accurate reading.
Pressure:
Dia 20-150 mmHg
Sys: 50-280 mmHg
Blood pressure measurements determined with this device are
equivalent to those obtained by a trained observer using the cuff
/stethoscope auscultation method, within the limits prescribed by
the American National Standard, Manual, electronic or automated
sphygmomanometers.
Operating:
10°C (50°F) to 40°C (104°F) Less than 95% RH non-condensing.
Storage:
-20°C (-4°F) to 50°C (122°F) Less than 95% RH non-condensing.
The system might not meet its performance specifications if used or
stored outside the specified temperature and humidity ranges listed
above.
External power supply, use only SunTech part number 19-0012-00.
Input: 100-240 VAC @ 50-60 Hz. Output : 9VDC @ 5A IEC 320
type input connector.
The accuracy of cuff -pressure transducers/indicators should be
verified annually.
software overpressure algorithm to limit cuff pressure to less than
300 mmHg (+20/-10mmHg). Independent hardware timing
circuit and redundant software timer algorithm to limit the duration
of a blood pressure cycle to less than 180 seconds.
Size: 22.8 cm x 15.9 cm x 9.7 cm (9.0" x 6.3'' x 4.1'')
Weight: 1.2 Kg (42.6 oz)
Heart Rate: 40-200 BPM
(beats per minute)