Maintenance And Cleaning - SunTech Tango+ Guia Del Usuario

Ocultar thumbs Ver también para Tango+:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

Maintenance and Cleaning

Preventative Maintenance
Establishing simple care guidelines helps protect the performance and life of your
Tango+. On a routine basis, you should inspect the device, cables and pneumatic
hoses for cracks, fraying or kinks and immediately replace any damaged parts.
The monitor performs a range of system and software checks during normal
operation, reporting to the user the operational status via the Status Display. Once
the monitor is on and the Main Measurement View is displayed, ensure that the
monitor's status display indicates the correct measurement mode. Also, verify that the
cuff pressure is 0 mmHg when no cuff is attached. DO NOT use the monitor if it has
failed any of its diagnostic self tests or if it displays a greater than zero pressure with
no cuff attached.
The monitor does not contain any user serviceable parts and should only be opened
by an authorized service representative. DO NOT remove covers or break the
warranty seal as this will void the manufacturer's warranty.
Calibration
It is recommended that the Tango+ have its calibration checked annually to verify the
accuracy of the pressure transducers and indicators.
Equipment Required:
1. Calibrated electronic manometer or equivalent.
2. 500mL volume or the Orbit-K Adult Plus cuff wrapped around something
that will not break or crush (no glass).
3. Hand Inflation Bulb with bleed valve.
4. Tubing, Tee pieces, and miscellaneous connectors or you can order the
T-Tube Kit (SunTech Part # 98-0030-00).
Procedure:
1. When the Main Measurement View is displayed, press the SELECT button 2
times. This will bring up the MAIN MENU screen.
2. Using the UP or DOWN arrows, highlight MONITOR SET UP and press the
SELECT button.
3. Using the UP or DOWN arrows, highlight VERIFY CALIBRATION and press
the SELECT button. The monitor will close its bleed valves and will display
on its screen the pressure applied to the patient hose connector.
4. Verify the Tango+ calibration by manually inflating and checking the
manometer against the pressure reading on the Tango+ display. The
Display should be within ±2mmHg of the pressure value on the mercury
manometer through a range of pressures from 0-300mmHg. If not, contact
SunTech Medical about calibration.
5. Once the calibration has been completed, press the SELECT button to exit
the calibration screen. Then press the UP or DOWN arrows to select EXIT
twice and return to the Main Measurement View.