Measurement Setup
Use to modify the blood pressure measurement settings.
Measurement
Mode:
Interval:
Initial Inflate:
Max Inflate:
Deflate Rate:
Beeper:
Stat Mode Key: Choose the measurement made by the monitor when in Stat
Choose the monitor measurement: BP or SYS.
BP includes both systolic and diastolic measurements and
deflates at the specified deflate rate (see Deflate Rate below).
SYS is a systolic measurement only and deflates at the AUTO
deflate rate, resulting in faster measurement times.
Set to MAN when interfacing a stress system with Tango+, so the
stress system can prompt Tango+ for a measurement. Also used
if you want to initiate each measurement using the START/STOP
key.
Otherwise, select a time interval (in minutes) between BP
measurements. The time intervals are 1:00, 1:30, 2:00, 2:30,
3:00, 4:00, 5:00, 10:00, or 20:00. If a specific time interval
is chosen, you must press the START/STOP key for Tango+ to
take the first measurement. The Tango+ will then automatically
initiate BP measurements at the specified interval.
Choose the pressure to which the cuff will inflate on the first
measurement of a study. Select a setting from 120 to 280 mmHg
(increments of 10).
Choose the maximum pressure allowed in the cuff during a
measurement. Select a setting from 120 to 280 mmHg
(increments of 10).
Choose AUTO for normal deflation of the cuff pressure.
Otherwise, enter a rate for special situations, such as standards
testing. Enter a rate from 3 to 8 mmHg/sec.
Choose when the Tango+ will make a sound indicating when the
measurement is started or completed.
BOTH: Beeps at the beginning and end of a measurement.
NONE: The system will not beep.
START: Beeps at the beginning of a measurement.
FINISH: Beeps at the end of a measurement.
mode:
BP: systolic and diastolic measurements
SYS: systolic measurements only
All measurements are made with the AUTO deflate rate.