Operational Functions - Tuthill TT10P Manual De Operación Y Seguridad

Ocultar thumbs Ver también para TT10P:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Operational Functions

• Turns meter on.
• Displays accumulated total while pressed.
If the accumulated total is larger than
9999, the numbers will scroll across the
screen in groups of 4 digits.
When held for 3 seconds, it resets the current
total to zero. Also reverts to normal operating
mode when in CAL or FLSH mode.
When held for 3 seconds, it allows changes
to the calibration/fl uid factor displayed in
the bottom left corner. Repeated activation
will increment the number up to 19 and
then back to zero. When the desired number
is displayed, press button 2 to lock in the
new number and return to normal operation.
Please reference Table 1 for specifi cs.
When held for 3 seconds, the RATE icon is
displayed. The active fl ow rate will be dis-
played. Press button 4 again to resume
normal operation. The meter will continue
to count volume while fl ow rate is displayed.
NOTE
The meter display will automatically turn off after
30 seconds of inactivity to conserve batteries. It will
turn on when fl ow resumes or any button is pushed.
No data is lost during periods of inactivity.
Changing Units of Measure
The units of measure can be changed to ounces (OZ),
pints (PT), quarts (QT), gallons (GAL), or liters (LITER)
without recalibrating the meter.
1. Hold buttons 2 and 4 for three seconds. The display
will show the fi rmware version.
2. Press button 3 . The display will show the current unit.
3. To change units, press the ON button repeatedly until
the desired unit is displayed.
4. Press button 2 to lock in the selection. The display will
show the fi rmware version.
5. To return to normal operating mode press button 2
again.
6. The units selected will be displayed. Current or
accumulated total will be updated to refl ect the new
unit of measure.
All manuals and user guides at all-guides.com
Press Calibration/Fluid Type Factor (CAL)
The TT10 Series Meters provide 20 preset factors for
different fl uid types/viscosities.
Use of calibration factors is intended to provide the most
accurate metering solution for a specifi c fl uid. A CAL factor
of 0 is for use with the thinnest fl uids and increases for
higher viscosity fl uids.
1. Press and hold button 3 for 3 seconds. The current
2. Repeatedly pushing button 3 will increment the
3. After the desired fl uid factor has been selected, press
CAL Factors and viscosities shown in the table below are
for fl uids at 70°F / 21°C.
3
3
dispensed count value will disappear and only the CAL
icon and the current value will be displayed.
number up to 19 and then back to zero.
button 2 to lock in the fl uid factor and return the meter
to normal operating mode.
Cal
0
1
2
3
cSt
Cal
10
11
12
13
14
cSt
Table 1: CAL Factor Table
NOTE
Use of the CAL/Fluid factor is independent from
performing a calibration against a known volume.
The meter is calibrated at the factory, but
performing a fi eld calibration may further improve
meter accuracy.
4
5
6
7
8
9
15
16
17
18
19
Tabla de contenido
loading

Tabla de contenido