Battery Replacement - Tuthill TT10P Manual De Operación Y Seguridad

Ocultar thumbs Ver también para TT10P:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Fluid Calibration
Fluid calibration requires a container of a known volume
at least 5 gallons or larger that is compatible with the fl uid
being metered.
A larger volume may be used and may improve calibration
accuracy, but do not exceed 99.99 units of measure.
NOTE
Before performing a fl uid calibration, ensure you
have selected the closest matching CAL/Fluid factor
to ensure the best meter accuracy.
1. Press the 2 & 4 buttons at the same time and hold
for 3 seconds. The display will show the fi rmware
version.
2. Press button 3 to enter calibration mode. The unit of
measure will be displayed.
3. Press button 3 . The display will read "FILL".
4. Now dispense fl uid into your container. "FILL" will blink
on the display when fl ow is detected.
5. After dispensing fl uid, press the ON button. The left
digit of the display will blink. ("00.00")
6. Press button 4 to increment the digit to the amount
of fl uid dispensed (example: "05.00"). Press the ON
button to move to the right. If you make a mistake,
press button 2 to start back at the left-most digit.
7. After number is loaded, press the ON button again
to accept the entered volume.
8. Display now shows the confi gured CAL factor.
9. Press the ON button again and the fi rmware version
will be displayed.
10. Press the 2 button again to return to normal
operating mode.
NOTE
If the value entered is out of an acceptable range,
the display will read "Err0" and the meter will revert
to the previous settings. See Troubleshooting Guide
for more information.
TIP
If you are metering in ounces, pints, or quarts, it
may be easier to switch the meter to a larger unit of
measure such as gallons or liters for the calibration
process and then change back to the desired unit of
measure.
All manuals and user guides at all-guides.com
Use of Flush mode allows you to fl ush the meter without
adding volume to the meter totals. This is useful when
fl ushing the meter for storage or switching between
different liquids and you want to maintain accurate totals.
1. Hold buttons 3 and 4 for 3 seconds. The Meter will
2. Press button 2 to return to normal operating mode.
1. Press and hold buttons 2 and 4 for 3 seconds.
2. Press the ON button to display the accumulated total.
3. Hold buttons ON and 4 together for 5 seconds to
4. Press the 2 button twice to return to normal operating
You can check the internal calibration factor for the meter.
The displayed value is the number of sensor counts per
selected unit of measure and CAL/Fluid factor. The factory
may ask you for this value if troubleshooting the Meter.
1. Hold buttons 3 and 4 for 3 seconds. The Meter will
2. Press and hold buttons ON and 3 together. The
3. Press button 2 to return to normal operating mode.
Batteries should be replaced when the LOW battery
icon begins to blink on the display. The meter totals,
confi guration, and calibration are preserved while the
batteries are being replaced, so there is no need to
recalibrate afterwards.
4
4
Flush Mode (FLSH)
display "FLSH".
Reset Accumulated Total
The current fi rmware version will be displayed.
If over 9999 units, the accumulated total will scroll
across the display.
reset the accumulated total.
mode.
K-Factor Check
display "FLSH".
number of counts per unit of measure will be displayed.

Battery Replacement

WARNING!
When replacing batteries, make the area safe of
explosive vapors, or move the meter electronics to
a safe area before replacing them.
Tabla de contenido
loading

Tabla de contenido