Operational Functions

Turns meter on.
Displays accumulated total while pressed.
If the accumulated total is larger than
9999, the numbers will scroll across the
screen in groups of 4 digits.
When held for 3 seconds, it resets the current
total to zero. Also reverts to normal operating
mode when in CAL or FLSH mode.
When held for 3 seconds, it allows changes
to the calibration/fluid factor displayed in
the bottom left corner. Repeated activation
will increment the number up to 19 and
then back to zero. When the desired number
is displayed, press button 2 to lock in the
new number and return to normal operation.
Please reference Table 1 for specifics.
When held for 3 seconds, the RATE icon is
displayed. The active flow rate will be dis-
played. Press button 4 again to resume
normal operation. The meter will continue
to count volume while flow rate is displayed.
NOTE
The meter display will automatically turn off after
30 seconds of inactivity to conserve batteries. It will
turn on when flow resumes or any button is pushed.
No data is lost during periods of inactivity.
Changing Units of Measure
The units of measure can be changed to ounces (OZ),
pints (PT), quarts (QT), gallons (GAL), or liters (LITER)
without recalibrating the meter.
1. Hold buttons 2 and 4 for three seconds. The display
will show the firmware version.
2. Press button 3 . The display will show the current unit.
3. To change units, press the ON button repeatedly until
the desired unit is displayed.
4. Press button 2 to lock in the selection. The display will
show the firmware version.
5. To return to normal operating mode press button 2
again.
6. The units selected will be displayed. Current or
accumulated total will be updated to reflect the new
unit of measure.
Press Calibration/Fluid Type Factor (CAL)
The TT10 Series Meters provide 20 preset factors for
different fluid types/viscosities.
Use of calibration factors is intended to provide the most
accurate metering solution for a specific fluid. A CAL factor
of 0 is for use with the thinnest fluids and increases for
higher viscosity fluids.
1. Press and hold button 3 for 3 seconds. The current
dispensed count value will disappear and only the CAL
icon and the current value will be displayed.
2. Repeatedly pushing button 3 will increment the
number up to 19 and then back to zero.
3. After the desired fluid factor has been selected, press
button 2 to lock in the fluid factor and return the meter
to normal operating mode.
CAL Factors and viscosities shown in the table below are
for fluids at 70°F / 21°C.
Cal
0
1
2
cSt
Cal
10
11
12
cSt
Table 1: CAL Factor Table
NOTE
Use of the CAL/Fluid factor is independent from
performing a calibration against a known volume.
The meter is calibrated at the factory, but
performing a field calibration may further improve
meter accuracy.